Date | Event | Location | Comment |
---|---|---|---|
Jan. 14, 2022 | LAUNCHED | Atlantic | |
Jan. 10, 2022 | LOADED | Atlantic | |
Dec. 7, 2021 | SHIPED_TO_PORT | Bermuda | |
Dec. 5, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Dec. 1, 2021 | PASSED_DOCKTEST | WHOI | |
Nov. 24, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Nov. 4, 2021 | PASSED_INCOMING | WHOI | Reference spectrum of SUNA updated. |
Nov. 1, 2021 | RECIEVED | WHOI | |
July 21, 2021 | RETURNED | Bellevue, WA | RMA: 1005513728. Failed pH sensor. |
April 27, 2021 | MODEM_SUSPENDED | WHOI | |
Jan. 12, 2021 | MODEM_ACTIVATED | WHOI | |
Jan. 1, 2021 | RECIEVED | WHOI | Check date |
Date | Event | Location | Comment |
---|---|---|---|
May 5, 2021 | LAUNCHED | N Atlantic | |
April 19, 2021 | SHIPED_TO_PORT | UK | |
Jan. 12, 2021 | MODEM_ACTIVATED | WHOI | |
Jan. 1, 2021 | RECIEVED | WHOI | Check date |
Date | Event | Location | Comment |
---|---|---|---|
May 5, 2021 | LAUNCHED | N Atlantic | |
March 19, 2021 | SHIPED_TO_PORT | UK | |
Jan. 1, 2021 | RECIEVED | WHOI | Check date |
Dec. 21, 2020 | MODEM_ACTIVATED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Sept. 30, 2021 | RETURNED | Bellevue, WA | Bumps and scratches in hull face seal |
Sept. 29, 2021 | RECIEVED | UW | |
Sept. 15, 2021 | RETURNED | Bellevue, WA | Scratches in hull seal face, swapping with hull from F1344 |
Aug. 16, 2021 | PASSED_PRESSURE_TEST | UW | |
June 25, 2021 | PASSED_DOCKTEST | UW | |
June 24, 2021 | PASSED_INCOMING | UW | |
May 12, 2021 | RECIEVED | UW |
Date | Event | Location | Comment |
---|---|---|---|
Sept. 15, 2021 | RETURNED | Bellevue, WA | Exposed metal in end cap, not anodized. Part will have to be replaced and float reassembled. |
Aug. 16, 2021 | PASSED_PRESSURE_TEST | UW | |
June 25, 2021 | PASSED_DOCKTEST | UW | |
June 24, 2021 | PASSED_INCOMING | UW | |
May 12, 2021 | RECIEVED | UW |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 18, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | pH reading slightly lower, but OK |
Sept. 30, 2021 | PASSED_DOCKTEST | WHOI | |
Sept. 28, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Sept. 17, 2021 | OTHER | WHOI | PASSED_DOCKTEST, reset for internal inspection |
Sept. 1, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Aug. 30, 2021 | PASSED_INCOMING | WHOI | 1 shock sticker triggered, pH is out of spec, SUNA unable to get sample. |
Aug. 27, 2021 | RECIEVED | WHOI | |
June 17, 2021 | RETURNED | Bellevue, WA | RMA: 1005513498. Returned for repair, failed vacuum test. Likely dropped during shipping |
June 11, 2021 | FAILED_INCOMING | WHOI | Failed tests |
June 1, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Aug. 12, 2021 | LAUNCHED | N Atlantic | |
July 8, 2021 | PASSED_DOCKTEST | WHOI | |
June 11, 2021 | PASSED_INCOMING | WHOI | Failed pH tests |
June 1, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 16, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 11, 2021 | OTHER | WHOI | Removed 10g of ballast. Still vertically stable in tall tank. |
Sept. 30, 2021 | PASSED_DOCKTEST | WHOI | |
Sept. 28, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Sept. 17, 2021 | OTHER | WHOI | PASSED_DOCKTEST, reset for internal inspection |
Sept. 1, 2021 | PASSED_PRESSURE_TEST | WHOI | |
July 1, 2021 | PASSED_INCOMING | WHOI | |
June 21, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 16, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | pH reading slightly lower, but OK |
Oct. 11, 2021 | OTHER | WHOI | Removed 10g of ballast. Still vertically stable in tall tank. |
Sept. 30, 2021 | PASSED_DOCKTEST | WHOI | |
Sept. 29, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Sept. 17, 2021 | OTHER | WHOI | PASSED_DOCKTEST, reset for internal inspection |
Sept. 9, 2021 | PASSED_INCOMING | WHOI | pH is out of spec, reference spectrum of SUNA updated. |
Sept. 1, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Aug. 27, 2021 | RECIEVED | WHOI | |
July 2, 2021 | RETURNED | Bellevue, WA | RMA: 1005513612. Returned for repair, failed vacuum test. Likely dropped during shipping |
June 29, 2021 | FAILED_INCOMING | WHOI | bad vacuum |
June 21, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Oct. 7, 2021 | RETURNED | Bellevue, WA | RMA: 1005514311. Salt internal. |
Sept. 29, 2021 | FAILED_INTERNAL_INSPECTION | WHOI | Found salt internal |
Sept. 20, 2021 | OTHER | WHOI | PASSED_DOCKTEST, reset for internal inspection |
Sept. 10, 2021 | PASSED_INCOMING | WHOI | Reference spectrum of SUNA updated. |
Sept. 3, 2021 | FAILED_INCOMING | WHOI | SUNA is reading high, SUNA is not getting measurement from CTD. |
Sept. 1, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Oct. 28, 2021 | RETURNED | Bellevue, WA | RMA: 1005514500. CTD communication error. |
Oct. 27, 2021 | FAILED_PORT_TEST | St. Petersburg, FL | Reversed capacitor |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 11, 2021 | OTHER | WHOI | Removed 10g of ballast. Still vertically stable in tall tank. |
Oct. 11, 2021 | PASSED_DOCKTEST | WHOI | |
Oct. 6, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Sept. 29, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Sept. 23, 2021 | PASSED_INCOMING | WHOI | |
Sept. 22, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 21, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 11, 2021 | PASSED_DOCKTEST | WHOI | |
Oct. 6, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Sept. 29, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Sept. 23, 2021 | PASSED_INCOMING | WHOI | |
Sept. 22, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 18, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 12, 2021 | PASSED_DOCKTEST | WHOI | |
Oct. 8, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Sept. 30, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | Missing white wire spacer, loose nut in hull. |
Sept. 23, 2021 | PASSED_INCOMING | WHOI | pH out of spec |
Sept. 22, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Oct. 20, 2021 | RETURNED | Bellevue, WA | RMA: 1005514397. Failed pH |
Oct. 13, 2021 | FAILED_FLOW_THROUGH | WHOI | pH sensor failed. |
Oct. 11, 2021 | FAILED_DOCKTEST | WHOI | pH sensor failed. |
Oct. 6, 2021 | PASSED_PRESSURE_TEST | WHOI | Salt loop came off. |
Oct. 1, 2021 | RECIEVED | WHOI | |
Oct. 1, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Oct. 1, 2021 | PASSED_INCOMING | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Nov. 12, 2021 | RETURNED | Bellevue, WA | RMA: 1005514595. Reversed Capacitor |
Nov. 4, 2021 | FAILED_INTERNAL_INSPECTION | WHOI | Reversed capacitor |
Oct. 1, 2021 | FAILED_INCOMING | WHOI | Bad vac, MCOMS channel 2 reading high, pH out of spec. Replaced O-ring and re-pulled vacuum. Currently monitoring. |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 23, 2022 | LAUNCHED | Atlantic | |
Dec. 28, 2021 | LOADED | Atlantic | |
Oct. 27, 2021 | PASSED_PORT_TEST | St. Petersburg, FL | |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 12, 2021 | PASSED_DOCKTEST | WHOI | |
Oct. 8, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Oct. 8, 2021 | PASSED_INCOMING | WHOI | SUNA was not configured. Set params and cal file. Retrieved cal file from instrument (not supplied on CD). |
Oct. 4, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Oct. 1, 2021 | RECIEVED | WHOI | |
Oct. 1, 2021 | FAILED_INCOMING | WHOI | SUNA cal file is missing, error when trying to recal. |
Date | Event | Location | Comment |
---|---|---|---|
Oct. 28, 2021 | RETURNED | Bellevue, WA | RMA: 1005514500. CTD communication error. |
Oct. 27, 2021 | FAILED_PORT_TEST | St. Petersburg, FL | Reversed capacitor |
Oct. 21, 2021 | SHIPED_TO_PORT | St. Petersburg, FL | |
Oct. 13, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Oct. 12, 2021 | PASSED_DOCKTEST | WHOI | |
Oct. 8, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Oct. 5, 2021 | PASSED_INCOMING | WHOI | MCOMS channel 2 seems to be reading high, channel 3 reading low. But this is similar to how others have behaved. |
Oct. 5, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Oct. 5, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Jan. 14, 2022 | LAUNCHED | Atlantic | |
Jan. 10, 2022 | LOADED | Atlantic | |
Dec. 7, 2021 | SHIPED_TO_PORT | Bermuda | |
Dec. 5, 2021 | PASSED_FLOW_THROUGH | WHOI | Salinity reading slightly fresh compared to 1358 and 1200. |
Dec. 1, 2021 | PASSED_DOCKTEST | WHOI | |
Nov. 24, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Oct. 18, 2021 | PASSED_INCOMING | WHOI | pH out of spec. |
Oct. 5, 2021 | RECIEVED | WHOI |
Date | Event | Location | Comment |
---|---|---|---|
Dec. 5, 2021 | PASSED_FLOW_THROUGH | WHOI | |
Dec. 1, 2021 | PASSED_DOCKTEST | WHOI | |
Nov. 24, 2021 | PASSED_PRESSURE_TEST | WHOI | |
Nov. 4, 2021 | PASSED_INCOMING | WHOI | |
Nov. 4, 2021 | PASSED_INTERNAL_INSPECTION | WHOI | |
Nov. 1, 2021 | RECIEVED | WHOI |